LB 470RID Level
6 Operation of the Software
56925-1BA2 Rev.02, 05/2021
61
6.3.3
Event Reports
Events are displayed in the standard display and in the submenus and windows as
a symbol. All events are displayed on the main screen. A specific “D” (for detector)
indicates that a detector has an event, the prefix “M” (for measuring unit) indicates
that there is an event in the LB 470RID transmitter. In the event of a detector fault,
the operating manual of the detector must be observed.
Only the event with the highest priority will be displayed. Refer to the menus Trans-
mitter Events (chapter 8) and Detector Event Log "Detector-Service" (chapter 7.3.2)
for additional information.
1
2
3
4
Event reporting in the standard display
Event message in the menu structure
Device concerned (detector / measuring instrument)
Button
<Confirm>
5
6
7
Error number
Title of event report
Button
<Close>
Fig. 30
Event reports (examples)
1.
Click on the icon (Fig. 30, item 1, item 2) to display detailed information
about the event.
2.
Click the button
<Acknowledge>
to confirm an event that requires a manual
confirmation.
The event description indicates the next event or reports no further events.
3.
Click
<Close>
to return to the submenu or to the standard display.
The icon disappears from the status information.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...