Chapter 15 Troubleshooting and correcting faults
Fig. 15-6
Left: USB stick with extended plug (operating position); Right: USB- stick with retracted plug
(position to transport and store the stick)
15.8
Circuit breaker and battery fuse
The Ikus battery pack is protected against excessive current.
There is an internal battery fuse that protects against excessive load current. From
outside it is unamenable and after triggering it has to be replaced by the service team.
A second fuse (the circuit breaker) protects against excessive charging current. The
circuit breaker is located on the connection panel and it is resettable after one-time
triggering (button Circuit breaker flips out) by the operator.
Activated circuit breaker or internal battery fuse in mains operation
WARNING
Never disconnect the Ikus from the mains when the circuit breaker or
the internal battery fuse are activated. This will cause the driving unit to
stop immediately.
INSTRUCTION
1.
Ensure that the Ikus is connected to the mains.
2.
Check if the circuit breaker was triggered. Push the button Circuit breaker
back in place to produce power supply again. Important: Only press the button
briefly. Never keep the button pressed for a longer period, because otherwise
the retriggering of the circuit breaker would not be detected.
3.
If the circuit breaker is reactivated, do not press the button in again. Never
keep the button pressed for a longer period of time. If possible, support the
patient with a replacement Ikus (see section 15.4: Connecting the patient to a
replacement Ikus, page 202).
HOTLINE
Notify Berlin Heart! 866.249.0128
Activated circuit breaker or internal battery fuse in battery operation mode
WARNING
The Ikus stops immediately. The blood pumps are no longer being
driven.
Whenever the Ikus is running in battery operation, the patient must
always be accompanied by a person trained to use the manual pump.
NOTICE
If the circuit breaker or internal battery fuse is activated in battery
operation, the Ikus generates an acoustic alarm.
1000721x09 Revision 8
208
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...