CAUTION
Chapter 13 Weaning and Explantation for BTR and BTT
Always follow the above sequence of operations. Always use the key
switch to switch off the Ikus.
Do not switch the Ikus off unless the batteries are fully charged. To do
this leave the Ikus switched on until all yellow LEDs light up, then switch
the Ikus off using the key switch.
INSTRUCTION
1.
Put the patient on cardiopulmonary bypass.
2.
Disconnect the driving tubes and connect both tank units to the Ikus.
3.
Leave the Ikus running with the tank units until the patient is stable on CPB
and the blood pumps have been explanted.
4.
Next in the monitor program, select the Drive OFF option and press <Enter> to
confirm (see Fig. 13-2, page 169).
5.
Respond to the prompt in the dialog window by pressing the <X> key or the
<1> key. The system stops operation immediately and writes an operating log.
6.
Disconnect the driving tube(s) from the connector(s). To do so, take hold of the
release sleeve and pull this out of the connector.
7.
Use the seal plugs to seal the driving tube connectors.
8.
Wait until the log has been completed. When the message Switch drive off
with main switch! appears, press <F10> to shut down the monitor program.
Confirm by pressing the <X> key or the <1> key.
9.
Select 3. End (<3>, see Fig. 13-3, page 169) in the start menu and switch off
the laptop.
10.
Switch the Ikus off, provided that the batteries are fully charged. To do so, turn
the key switch to [0] position.
Fig. 13-2
Drive OFF
Fig. 13-3
Start menu
1000721x09 Revision 8
169
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...