Function
Batteries charge most quickly at room temperature, between 19 °C and 23 °C. Always
charge batteries fully whenever possible.
Connecting the battery charging unit
Charging the battery and checking the charge level
illuminated green
The battery charging unit is connected
to the mains, the battery mount is
ready for operation.
chaser light
The battery is charging.
green, steadily illumi-
nated
The battery is fully charged.
orange, flashing
The battery is defective and may not be
used further.
orange, flashing
The battery mount is defective. The
battery charging unit must not be used
again.
orange, steadily illu-
minated
The battery mount is too warm. Place
the battery charging unit in a cooler
location.
1.
Connect the power supply unit to the battery
charging unit.
2.
Check the readiness for operation. The sta-
tus indicator must be illuminated green on
both sides.
1.
Insert the battery into a free battery mount
from above.
2.
Check the charge activity indicator. A green
chaser light shows the charging process.
3.
If all charge activity indicators are illumi-
nated green, the battery is fully charged.
4.
Removal: Pull the battery upwards.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...