1015002x00
125
Replacing the Driving System
en
9.3
From EXCOR Active to Ikus
9.3.1
Preparing the Ikus
9.3.2
Replacing the Driving Unit
1.
Switch on and configure the Ikus, see the Ikus instructions for use.
2.
Transfer the EXCOR Active driving parameters to the Ikus. The following
parameters are transferred (for each supported side in BVAD mode):
•
Operating mode
•
Cannula sizes
•
Size of the blood pump
•
Rate
•
Driving pressure in systole
•
Driving pressure in diastole
•
Synchronization
type (in BVAD operating mode)
3.
Initially, leave the parameter
Systolic duration
on the Ikus to the default
value of 40%.
4.
Position the driving unit on the Ikus so as to allow fast and secure reconnec-
tion of the driving tube(s).
1.
Lay the patient on his or her back. Perform the following steps (for each sup-
ported side in BVAD mode):
2.
Remove the cable wrap from the driving tube.
3.
Disconnect the flow sensor from the cannula and the driving unit. Press
A
UDIO
PAUSED
.
4.
Disconnect the tank unit from Ikus.
5.
Switch the driving tube from EXCOR Active to Ikus.
6.
Check filling and emptying of the blood pump and adjust as needed.
7.
EXCOR Active driving unit: Stop blood pump operation. Use the seal plugs to
close the pneumatic connections on EXCOR Active.
8.
Allow EXCOR Active to run for a time, in parallel with Ikus. Switching back is
easy to do if necessary. After that, switch off the driving unit, see section 5.6:
Switching Off the Driving Unit, page 50.
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...