4.8 Movement of the OR Table (on casters)
NOTICE
Certain procedures require that the table be moved after anesthe‐
tizing the patient. If this is required, the following steps must be
followed:
1. Be sure that the patient is properly secured to the table
with straps.
2. Center (slide) the patient over the main lift column
3. Lower the table to its lowest point.
4. Check the floor for and remove objects that may inter-
fere with moving the table, such as power cords or table
accessories.
5. Have at least one person at each end of the table
6. Unlock the table.
7. When moving the table, the people at each end must
use a lateral force in the direction of motion; do not lift
up on a section or push down on a section.
8. Immediately lock the table when the new location is
reached.
The InstaDrive™ option is available on the OPERON Mobile tables. The in-
tended use of the function is to aid in the transport or movement of OR
table from room to room. The OR table with or without the InstaDrive™
option is not approved for patient transport. However, there will be cer-
tain procedures that require that the table be placed on the castor or
wheels and repositioned after anesthetizing the patient. In such cases,
the conditions listed in section 4.9.1 must be met.
CAUTION
Two persons are required to move the OR table!
Moving the OR table:
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Operation
35