3.7 Checks before initial start-up
NOTICE
The hydraulic fluid can contract/expand greatly at storage tempe-
ratures below 50°F and above 100°F, adversely affecting the stor‐
ed sections. This situation resolves itself approximately 24 hours
after the OR table is installed in a room and reaches room tem‐
perature.
1. Make sure that all accessories and attachments are removed from
the packing material.
2. Visually inspect the OR table to ensure that all straps, spacers and
other packing materials have been removed.
3. Ensure that the supply voltage is 110/120 V AC 60 Hz or 230 V AC 50
Hz.
4. Check all motion limits (“Motion range limits” section) and all func-
tions (“Operation” section) of the OR table.
5. Check the function of the auxiliary hand pendant (“Auxiliary hand
pendant” section) and the foot pump ("Operating the foot pump”
section).
6. Visually check for leaking hydraulic fluid, especially under the table,
at the column head covers and under the side panels. If there is a
leak, contact BERCHTOLD Technical Service at 800-243-5135, Option
2.
Operator Manual Surgical Mobile Operating Table Models D860, D850, D 830, D820, D770, D760
Installation and initial start-up