F1.2: Calibration mode selection, with 2 options
F1.2=1 fixed point calibrating
F1.2=2 arbitrary point calibrating
Note: If the version number is less than REV2.0, choose F1.2=2 only
F1.3: Baud rate selection, with total 5 options
F1.3=1 B =600
F1.3=2 B =1200
F1.3=3 B =2400
F1.3=4 B =4800
F1.3=5 B =9600
F1.4: Auto shut-off function selection, with 3 options
F1.4=0 disabled
F1.4=3 auto shut-off in 3 minutes after stopping operating
F1.4=5 auto shut-off in 5 minutes after stopping operating
F1.5: Key sound selection, with 2 options
F1.5=0 disabled
F1.5=1 allowed
F1.6: Setup value selection from F1.1 to F1.5, with 2 options
F1.6=1 select setup value of customer
F1.6=2 select the default value , see the following table:
Note: After the F1.6 is selected and press [ENTER] to confirm, the indicator
will display “ F1” again. To re-enter by pressing [ENTER] and press [EXIT]
to withdraw. And also to press [
] to re-select main directories.
5.2.2 Subdirectories F2
F2.1: Turn-on zero setting range selection, with 4 options
F2.1=0 disabled
F2.1=5
±
5%F•S
F2.1=10
±
10%F•S
F2.1=20
±
20%F•S