19
BENTHAM | TLS120Xe Product Guide v1.3
Light. Measurement. Excellence.
The USB IDs of the TLS120Xe are VID 04D8, PID 1705.
Upon finishing the boot up sequence, the user can control the unit remotely.
In the following example the TLS120Xe is set to remote mode, the target
wavelength is set to 500nm, the target filter set to the appropriate filter for
500nm and the monochromator moved to the target wavelength:
:SYST:REM
:MONO 500
:MONO:FILT:WAVE 500
:MONO:MOVE
Please see the TLS120Xe communications manual for
further information.
To revert to local mode after computer control, hold the output button for 3
seconds.
Figure 7: Display shown when initiating remote mode