Measuring Adapter BENNING MA 2-16

 

 

 

1.  Application 

With the 

BENNING MA 2-16 

measuring adapter you can test 

devices equipped with a 5-pin 16 A/ 6h CEE plug quickly and 
efficiently.  To  do  this  connect  the  adapter  to  a  tester  for 
testing 

portable 

devices 

in 

compliance 

with 

DIN VDE 0701-0702  or  for  testing  in  compliance  with 
DIN EN 62353 (VDE 0751-1)  or  DIN EN 60974-4 (VDE 0544-4), 
that has only a shock-proof socket for connecting to the test 
sample  and  that  can  measure  protective  conductor  current 
directly. 

The adapter is suitable for connection to the appliance testers  

BENNING ST 755

 and 

BENNING ST 760

The  following  tests  can  be  carried  out  with  the  measuring 
adapter on appliances with CEE plugs: 

  Test of the continuity of the protective conductor system 

  Test of insulation 

  Measurement of the protective conductor current with 

the method:  
direct measurement method/ differential and alternative 
leakage current 

  Functional test 

 

Attention! 
Restriction of the application 

The 

BENNING MA 2-16

 measuring adapter may only 

be used for testing purposes in combination with tester 
types 

BENNING ST 755

 and 

BENNING ST 760

. Any 

other use of the adapter, in particular for permanent 
connection  of  three-phase  appliances  to  a  three-
phase  mains,  is  not  permissible  under  any 
circumstances! 

The  maximum  permissible  thermal  continuous 
current in all cases is 16 A AC per phase.

 

2.  Safety features and measures 

The  measuring  adapter  is  constructed  and  tested  in 
compliance 

with 

the 

safety 

regulations 

IEC 61010-1/ EN 61010-1/ VDE 0411-1. The safety of users 
and  the  device  is  guaranteed  if  it  is  used  for  its  intended 
purpose. 

The device may only be connected to a TN, TT or IT power 
supply  of  max.  240 V/ 400 V,  which  complies  with  the 
applicable  safety  regulations  (e. g. IEC 60346, VDE 0100)  and  is 
secured with a maximum nominal current of 16 A. 

Before use please read this instruction manual carefully and 
completely  and  in  particular  the  instruction  manual  of  the 
tester with  which you intend to use this measuring adapter. 
Observe and follow all points included in the instructions. 

The adapter may only be used for testing appliances with a 
5-pin  CEE  plug  and  a  maximum  current  consumption  of 
3 x 16 A (AC-1). 

The measuring adapter must not be used: 

  with the housing open 

  if there are any signs of external damage 

  if the CEE socket or connection cable are damaged 

  after severe overuse, i. e. exceedance of the load limits 

in the technical specifications 

  after storage over a longer period of time under 

unfavourable conditions  
(e. g. humidity, dust, temperature) 

Meanings of the symbols on the device 

 

Attention: Observe documentation! 

This symbol means that the instructions in the 
instruction manual must be observed in order to 
prevent danger.

 

 

CAT II 

Device in the measuring category II 

 

Earth (voltage to earth)

 

 

This symbol on the measuring adapter means 
that the measuring adapter conforms to the EU 
directives.  

 

The device must not be disposed of with 
household waste. You will find more information 
about the WEEE symbol on the Internet under 
the search term WEEE. 

3.  Connecting the BENNING MA 2-16 

Before  you  can  start  testing,  you  have  to  connect  the 

BENNING MA 2-16 

to your tester and the mains. To do this 

proceed as follows: 

  Connect  the  mains  cable  of  your  tester  to  the  shock-

proof socket on the 

BENNING MA 2-16 

marked “mains 

connection 

tester“. 

  Connect the protective contact line of the   

BENNING MA 2-16

  to  the  test/ mains  socket  of  your 

tester.  The  connection  on  the 

BENNING MA 2-16

  is 

marked 

“for test/ mains socket tester”. 

  Connect  the  CEE  connection  lead  of  the 

BENNING 

MA 2-16

  to  a  suitable  CEE  socket  16 A/ 6h  in  your 

electrical 

installation. 

The 

connection 

on 

the 

BENNING MA 2-16

 

is 

marked 

“3~/ N/ PE, 400 V, 50-60 Hz, 16 A”. From this time onwards 
the 

BENNING MA 2-16

 and your tester will be supplied 

with electrical power. 

  Finally  connect  your  test  sample  to  the  CEE socket  of 

the 

BENNING MA 2-16

.  The  socket  is  marked 

“Test/ mains socket test sample 16 A”. 

You can now start the tests. (s. Chapter.4) 

4.  Testing with the BENNING MA 2-16 

All  tests  that  you  can  carry  out  with  your  tester  on  devices 
with  a  shock-proof  plug  and  that  are  supported  by  the 

BENNING MA 2-16 

adapter

 

can  now  be  carried  out  in  an 

identical manner on devices with a CEE plug. Proceed here 
as indicated in the operating manual of your tester. 

The following special features of the 
BENNING MA 2-16 adapter must be taken into 
account during testing with the tester. 

  In  the  case  of  an  insulation  or  alternative  leakage 

current test, the three phase connections L1, L2, L3 and 
N  of  the  test  sample  are  short-circuited  in  the  adapter 
and then the tests carried out. 

  In the case of a protective conductor resistance test, the 

value of the measured protective conductor resistance 
is  increased  by  the  proportion  that  is  caused  by  the 
protective conductor system of the 

BENNING MA 2-16 

measuring adapter itself.  

 

 