03/ 2020
BENNING IT 105 CH
26
►
►
►
I
∆
N nominal fault current (F4):
Press the function key F4 to select the nominal fault current:
Available nominal fault currents (sinusoidal testing current) :
10 mA
30 mA
100 mA
300 mA
500 mA
½ I
∆
N
1 I
∆
N
5 I
∆
N
- Connect the measuring leads to the BENNING IT 105 as shown in figures 4 or 5 and apply them to the
test object.
- Press the TEST key
4
to start the measurement.
- The measured tripping time is shown on the main display
K
.
8.6.3 Tripping current RCD
I
- Select the desired function RCD
I
M
with the rotary switch
1
.
- The symbols of the function keys F1
A
to F4
D
are briefly shown on the digital display
2
. You can
use the function keys F1 to F4
3
to make the following settings:
F1
F2
F3
F4
0° / 180°
–
I
∆
N
0°/ 180° (F1):
0°:
Testing current with positive initial polarity
180°: Testing current with negative initial polarity
RCD type (F2):
sinusoidal test current
pulsating test current
selective (delayed) RCD
For testing selective RCDs, measurement starts after a time delay of 30 seconds.
►
►
►
I
∆
N nominal fault current (F4):
Press the function key F4 to select the nominal fault current:
Nominal fault currents at RCD type AC :
10 mA
►
30 mA
►
100 mA
►
300mA
►
500 mA
- Connect the measuring leads to the BENNING IT 105 as shown in figures 4 or 5 and apply them to the
test object.
- Press the TEST key
4
to start the measurement.
- The measured tripping time is shown on the main display
K
.