Maintenance
Regular Check
TMS60 Operator’s Manual
16 - 3
16.3
Regular Check
Perform a visual inspection before the equipment is first used every day. Verify that the
equipment meets the following requirements:
■
The housing and display screen are free from cracks or other damages.
■
All keys funtion properly.
■
Connectors are not loose, cracked, or bent and cables have no cuts, nicks, or
fraying.
■
ECG leadwires are securely connected with the equipment
■
Battery pack is installed and has sufficient charge.
After your equipment has been used for 6 to 12 months, or whenever your equipment is
repaired or upgraded, a thorough inspection should be performed by qualified service
personnel to ensure the reliability.
Follow these guidelines when inspecting the devices:
■
Make sure that the environment and power supply meet the requirements.
■
Inspect the devices and their accessories for mechanical damage.
■
Inspect all plugs, connectors, leadwires for damage, and make sure that their
insulation is in good condition.
■
Make sure that only specified accessories are applied.
■
Make sure that the alarm system functions correctly.
■
Make sure that the battery meet the performance requirements.
■
Make sure that the devices are in good working condition.
In case of any damage or abnormality, do not use the devices. Contact the hospital’s bio-
medical engineers or your service personnel immediately.
Summary of Contents for TMS60
Page 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Page 2: ......
Page 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Page 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Page 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Page 235: ...TMS60 Operator s Manual B 1 B EMC ...
Page 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Page 249: ......
Page 250: ...P N 046 007056 00 4 0 ...