11 - 6
BeneVision TM80 Telemetry Monitor Operator’s Manual
◆
The number of asterisk symbols before an event indicate alarm priority as
described in
11.3.3 Example Event Details Page
After selecting the desired event overview area, you can view the event details.
■
For manual events, you can view all the monitored waveforms 16 seconds before and
after the event trigger time and all the measurement numerics at the event trigger
time.
■
For alarm events, you can view event-related waveforms 16 seconds before and after
the event trigger time and all the measurement numerics at the event trigger time.
The following figure is for reference only.
The event trigger time is marked in a colored block along the timeline at the top of the
event details page. Different color blocks indicate different types of events:
■
Red: high priority alarm event
■
Yellow: medium priority alarm event
■
Cyan: low priority alarm event
■
Green: manual event
■
White: operation-related event
All the measurement numerics at the event trigger time are displayed at the bottom of the
event details page. You can view these numerics by swiping your finger at the bottom of
the main screen. The background color of numerics indicates the alarm priority.
■
Red: high priority alarm event
■
Yellow: medium priority alarm event
■
Cyan: low priority alarm event
Summary of Contents for TM80
Page 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Page 2: ......
Page 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Page 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Page 239: ......
Page 240: ...P N 046 012662 00 6 0...