Benetech GM510 Operation Manual Download Page 6

-08-

-09-

Reset and calibration

        

Before being sent out of factory, the device has 

been properly reset and calibrated with which 
consumers are not expected to operate.

Reset

Pressure difference shall read zero after turning on, if 
not, reset the device as the steps below: press the DIF 
key and DIFFERENTIAL PRESSURE shows on the LCD, 
meanwhile hold on pressing on the HOLD key for about 
3 seconds until the zero appears which indicates the 
reset is done, release the HOLD key and backlight key 
to return to normal measure mode

Calibration

In operation, if there is any error between the device 
and your standard, please calibrate it as the steps below:

Power on.
Hold on pressing the 

HOLD

 key and backlight key for 

     about 3 seconds until the reading

shows on

     LCD with its unit

that is inner value of the 

     device, which may be ignored by the users and 
     indicates it is under calibration mode and the 2 keys 
     said above could be released.

1

.

2

.

(1)
(2)

1.00

kPa

(3)

   

GM510/GM511

1

.

00kPa~10

.

00kPa

                                   

-1

.

00kPa~-10

.

00kPa

      GM520

1

.

00kPa~35

.

00kPa

                            

-1

.

00kPa~-35

.

00kPa

(4)

(5)

3.

(1)

(2)
(3)
(4)
(5)

(6)

(7)

Press on the 

DIF

 key to increase 1 kPa and hold on 

     pressing the 

DIF

 key for 2 seconds to increase the 

     value quicker. 

calibration range

calibration range

Press the 

HOLD

 key after selecting the desired value 

     for calibration and import the differential value into 
     the device.

Press the backlight to exit calibration mode and 

     return to normal measurement mode. If there is any 
     further calibration repeat the step4 and step 5.

Restore ex-factory setting

Users may restore the device to ex-factory settings if 
the calibration data confused due to mistake operation.

Holding on pressing the 

DIF

 key, 

HOLD

 key, backlight 

     key and 

REC

 key at one time

Power on the device
After full display and 0 appears release the 

REC

 key

At 00 appearing release the backlight key
At 000 appearing press the backlight key and 

REC

 

     key

At 0000 appearing release all the keys and complete 

    the operation

After restoring to ex-factory setting, refer to point 1 

    and point 2 above for reset and calibration if necessary.

10

.

Battery replacement

“    ”

flashes when the battery is insufficient, replace 

it with 4 1.5V AAA batteries.

Summary of Contents for GM510

Page 1: ......

Page 2: ...check the pack if there are following components and contact the local dealer if there is any missing in the pack or manual Brief There are following 11 selectable pressure units bar mbar kPa kgf cm mmHg cmH2O Ozf in psi inHg inH2O ftH2O This unit features data save and adjustment automatic turn off connecting with PC to read and export the data with a USB wire P 1 5V AAA battery Operation manual ...

Page 3: ...0 0 509 375 0 509 5 116 00 7 251 14 76 200 5 16 70 1 500 1500 0 150 1 529 1125 0 1528 5 348 00 21 755 44 29 601 5 50 10 Environmental temperature compensation Applicable to GM511 GM520 but not to GM510 03 02 Technical index Features 1 L 2 3 4 5 6 arge LCD Data hold With differential mode and record mode USB port Reset and data adjustment Low battery indication and automatic power off Measuring ran...

Page 4: ...pplication Definition of pressure Widely applied in measuring the pressure of fans resistance wind velocity oven pressure differential pressure of filter and monitoring the gas content in the air in process of combustion The pressure in this manual refers to the intensity of pressure the force applied on a square unit absolute pressure Difference between the reading and the absolute zero Positive ...

Page 5: ... in USB port allows connection with PC rror If there is any mistake reading appears it will be shown on the LCD as following Pressure beyond positive range ressure beyond negative range How to operate 1 P 2 3 4 1 2 ower supply Press ON OFF key to turn the unit on or off Difference Press the DIF key to display the value that is the difference between the current pressure and the pressure value when...

Page 6: ...kPa 1 00kPa 10 00kPa GM520 1 00kPa 35 00kPa 1 00kPa 35 00kPa 4 5 3 1 2 3 4 5 6 7 Press on the DIF key to increase 1 kPa and hold on pressing the DIF key for 2 seconds to increase the value quicker calibration range calibration range Press the HOLD key after selecting the desired value for calibration and import the differential value into the device Press the backlight to exit calibration mode and...

Page 7: ...roper transportation after purchasing and wrong storage Please refer to the guaranty card offered damage due to abuse failure to provide purchasing proof or unauthorized amendment on the purchasing invoice guaranty card 3 a b Claim We reserve the right to change the specification or instruction manual of the product without further notification Old battery dealing must be complied with the local l...

Reviews: