Chapter 9 Limitations
1. For In-vitro diagnostic use only (External use only).
2. BeneCheck Test Strip is designed for one-time use only. Do not reuse.
3. Do not perform the test while a mobile phone is used beside; prevent the
electronic signal interference.
9.1 Limitations of Total Cholesterol Test
BeneCheck Total Cholesterol Test Strips give accurate results when the following
limitations are observed :
●
Hematocrit (Hct) percentage may affect test result. If Hct levels less than 35%
may cause incorrect high measurement results and Hct levels greater than 50%
may cause incorrect low measurement results. Please consult your medical
professionals if you do not know your Hct level.
●
Use only fresh capillary finger whole blood. Do not use serum or plasma.