Modbus register assignment of the ISOMETER®
34
isoGEN523_D00320_04_M_XXEN/06.2018
RW
= Read/Write;
RO
= Read only;
WO
= Write only
8.1 Device-specific data type of the ISOMETER®
8.1.1
Device name
The data format of the device name is specified below.
8.1.2
Measured values
Each measured value is available as a channel and consists of 8 bytes (4 registers). The
first measured value register address is 1000. The structure of a channel is always iden-
tical. Content and number depend on the device. The structure of a channel is shown
with the example of channel 1:
9800
to 9809
RO
Device name
UNIT 16
(ASCII) -
refer to
---
---
9820
RO
Software
ID number
UINT 16
---
Software D
number
9821
RO
Software version
number
UINT 16
---
Software
version
9822
RO
Software version: Year UINT 16
9823
RO
Software version:
Month
UINT 16
9824
RO
Software version: Day UINT 16
9825
RO
Modbus driver version UINT 16
Word
0x00
0x01
0x02
0x03
-------------------
0x08
0x09
10 Words in total
Each Word contains two ASCII characters
1000
1001
1002
1003
HiByte
LoByte
HiByte
LoByte
HiByte
LoByte
HiByte
LoByte
Floating point value (Float)
Alarm type and
test type
(AT&T)
Range and
unit (R&U)
Channel description
Regpage
31ister
Property
Description
Format
Unit
Value range