background image

 

II-2. Installation

   

      

-

 Choose a location that is free from rapidly changing 

temperature conditions (away from a heating or cooling 

vent).    

-

  To assure proper ventilation, allow a minimum of 2 inches (5 

Cm) of the clearance between the rear and side of the unit 

and the adjacent wall.   

-

  Always ensure that the machine is installed on a level surface. 

-

  Before plugging the machine into the outlet, ensure that the 

electrical specifications (listed on the serial number label on 

the rear side of the unit) matches the electrical specifications 

of your outlet. 

-

 Always 

  

  - Avoid a direct ray and light.   

  

  - Avoid high humidity and dust air.   

  

  - Avoid cool areas (below 5 ).   

 

 

 

 

 

 

Summary of Contents for Incu-Shaker 10L

Page 1: ...Operations Manual Items H1010 and H1012 E...

Page 2: ...ex I General Information II 1 Introduction II 2 Installation II 3 Specifications II 4 Features III 1 Operation III 2 Platform Installation III 3 Error Messages III 4 Maintenance IV 1 Service and Conta...

Page 3: ...se check the Instrument and the accessories against the packing list when you first open the shipping carton If you find there is damage a discrepancy or missing items please contact your supplier imm...

Page 4: ...en the rear and side of the unit and the adjacent wall Always ensure that the machine is installed on a level surface Before plugging the machine into the outlet ensure that the electrical specificati...

Page 5: ...3 5 W x 29 2 D cm 9 5 W x 11 5 D in 2 Temperature range Ambient 5 C to 60 3 Temperature accuracy 0 1 at 37 4 Temperature uniformity 0 25 5 Speed range 30 300 RPM 6 Shaking Orbit 19mm 3 4 7 Circulation...

Page 6: ...2 D cm 9 5 W x 11 5 D in 2 Temperature range Ambient 15 C to 60 3 Temperature accuracy 0 1 at 37 4 Temperature uniformity 0 25 5 Speed range 30 300 RPM 6 Shaking Orbit 19mm 3 4 7 Circulation Fan Conv...

Page 7: ...l lid allows internal viewing without disturbance 4 Brushless DC motor and beltless drive system minimizes noise and vibration 5 Digital control panel displays temperature speed time 6 Optional access...

Page 8: ...nly If setting the machine to cool turn on the Cool Switch located on the front left side of the instrument DO NOT power on the Cool Switch if the temperature of the instrument is more than 5 C above...

Page 9: ...ed underneath the time speed display The set speed is displayed and will blink continuously It can be adjusted by using the UP and DOWN keys Once the desired speed is displayed press the SET key to co...

Page 10: ...Installation If you have ordered the H1010 MR MAGic clamp platform as an accessory with your Incu Shaker 10L platform may arrive installed on the shaker If not please follow the steps below for instal...

Page 11: ...Check the platform to ensure that nothing is obstructing it s movement Check the set speed to make sure that it is within the acceptable range 30 to 300rpm Solution Use your hand to move platform Whil...

Page 12: ...ost power during the run Solution Check that the power cord is properly installed and restart or check that the circuit has not been interrupted and restart 4 Display will not illuminate 1 Electrical...

Page 13: ...vent of a spillage Power down the unit remove the platform and dry the internal chamber with a soft cloth DO NOT attempt to operate the unit until the spillage had been properly dried IV 4 Service and...

Reviews: