BSH5000 Operation manual
Chapter 4 Operation Guide
—
9
—
The upper displayed temperature (Ex. 32.0) is the
current temperature measured by the External sensor.
When the temperature reaches to 40° and is held as constant
the decimal digit begins to flash.
Notes: Please allow 20 minutes before reading the
displayed actual value to ensure the calibration accuracy.
Press
“Mode” key
to
view the
temperature
value of
External
Sensor
. If the
displayed value is 39.6°C
.
Press “Mode” again, or wait for 6 seconds, then display
“
” , modify the displayed temperature to 39.6 by
pressing
or
. Then press
to confirm the input value.
d) Then the instrument will heat to 100
℃
automatically. Once 100
℃
has been
reached and held constant, the decimal point will begin to flash.
Notes: Once flashing, the user must allow 20 minutes before proceeding to the next
step. This will ensure calibration accuracy.
Press the “Mode” key to show the actual readout of
The external sensor (for ex. 101.5
℃
).
Press “Mode” again, or wait for 6 seconds, then display
“
” , modify the displayed temperature to 101.5 by
pressing
or
. Then press
to confirm the input value.
Note: Pressing “Mode” and “Start/Stop” simultaneously during the temperature
calibration indicates exiting the temperature calibration program. The