Benchmark Scientific BSH5000 Operating Manual Download Page 11

 

BSH5000 Operation manual

                              

      Chapter 4 Operation Guide 

 9 

 

The upper displayed temperature (Ex. 32.0) is the   
current temperature measured by the External sensor. 

 

 

 

 

 

   

When the temperature reaches to 40° and is held as constant   

    the decimal digit begins to flash.     

 

 

Notes: Please allow 20 minutes before reading the 
displayed actual value to ensure the calibration accuracy. 

 

 

Press

 “Mode” key

 

to

 

view the

 

temperature

 value of 

External

   

Sensor

.  If the 

displayed value is 39.6°C

 

 

 

Press “Mode” again, or wait for 6 seconds, then display   
“ 

  ” , modify the displayed temperature to 39.6 by   

pressing 

 or 

. Then press 

 to confirm the input value. 

 

 

 

d) Then the instrument will heat to 100

 automatically.  Once 100

 has been 

reached and held constant, the decimal point will begin to flash.     

 

Notes: Once flashing, the user must allow 20 minutes before proceeding to the next 

step.    This will ensure calibration accuracy. 

 

Press the “Mode” key to show the actual readout of   
The external sensor (for ex. 101.5

).   

 

Press “Mode” again, or wait for 6 seconds, then display   
“ 

  ” , modify the displayed temperature to 101.5 by   

pressing 

 or 

. Then press 

 to confirm the input value. 

 

 
Note: Pressing “Mode” and “Start/Stop” simultaneously during the temperature 

calibration indicates exiting the temperature calibration program.  The 

 

 

Summary of Contents for BSH5000

Page 1: ...Operation Manual BSH5000 myBlock Dry Bath Incubator...

Page 2: ...BOOKMARK NOT DEFINED CHAPTER 3 OVERVIEW 3 1 STRUCTURE OVERVIEW 3 2 KEYPADS 4 3 DISPLAY 4 CHAPTER 4 OPERATION 5 1 TEMPERATURE AND TIME SETTING ERROR BOOKMARK NOT DEFINED 2 ADVANCED TIME SETTING ERROR...

Page 3: ...mplification of DNA the pre denaturation of the electrophoresis blood serum coagulation and a variety of laboratory procedures Features Large digital display of time and temperature Extremely accurate...

Page 4: ...35 C Relative humidity 70 Power supply AC115V or 230V 50 60Hz 2 Technical Data Item No BSH5000 E Temperature range Ambient Room Temp 5 C 105 C Timer Max 99h59min Temperature accuracy 0 5 C Temperature...

Page 5: ...ew This chapter focuses on the introductions of the structure keypads and key functions of the instrument as well as preparatory work prior to initial operation 1 Structure overview Housing Power inle...

Page 6: ...key After the temperature and time setting is selected press this key to start the timer This key can also be pressed to stop the timer Down key Used in both time and temperature mode to decrease the...

Page 7: ...shown on the display are actual settings real time temperature time remaining The example shows an actual temp of 28 5 C and 35 min remaining c To adjust the set temperature use the and keys to selec...

Page 8: ...isplay will show OP 1 use the or key to select OP 2 The setting has now been changed and can be stored by pressing the Start Stop key 3 Temperature Calibration The temperature of the instrument has be...

Page 9: ...then press key at the same time and hold for 3 seconds it will enter into the temperature calibration interface see the right At this point displayed which indicates entering into temperature calibra...

Page 10: ...ressing or Then press Start Stop to confirm the input value Note Pressing Set and simultaneously during the temperature calibration indicates exiting the temperature calibration program The adjustment...

Page 11: ...displayed temperature to 39 6 by pressing or Then press to confirm the input value d Then the instrument will heat to 100 automatically Once 100 has been reached and held constant the decimal point wi...

Page 12: ...e drybath or directly into a sample tube b Press and hold Stop then simultaneously press and hold for 3 seconds the external indicator light goes on to indicate that it has entered into the temperatur...

Page 13: ...se On Off button broken Change button Others Contact the supplier 2 The actual and displayed temperatures are quite different Broken sensor or loose contact of the block Contact the supplier 3 OPEn di...

Page 14: ...al Annex 12 Annex Wiring Diagram for reference only L PE N Switch USB PT1000 CONN BOARD J3 J2 J1 AC 220V J4 Heat POWER BOARD J3 Hea tCtl J2 AC9V J2PT100 J4PT1000 USB 1 US B J1 AC 9V X1 He atCtl MAIN B...

Page 15: ...Memo...