Doc 0176160 rev. I
Page 19 of 19
Not subject to issue control when printed
Use of the test fixture is mandatory for all Paris Platform BTE during AUDIOmaster testing.
To use the fixture, place the instrument under test in the fixture according to picture.
Connect the programming cable to the programming adaptor and then attach the programming adaptor to the device.
Select test program in AUDIOmaster and press ‘Start’
During the test program execution the IAD calibration will be performed at the appropriate time. Another benefit from the fixture is that
the instrument is in a well defined position so the telecoil test can be performed without any interaction from the operator.
Test tubing is 20 mm effective length as for
all AUDIOmaster BTE testing
Place device in test chamber as shown
20mm
Testing