Annex of Report 

 

Manufacturer’s Declaration of the EUT 

(altogether 5 pages) 

Guidance and manufacturer’s declaration – electromagnetic emission –   
for all EQUIPMENT AND SYSTEMS 

 

Guidance and manufacturer’s declaration – electromagnetic emission 

The  SF-371  body  fat  scale    is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The 

customer or the user of SF-371 body fat scale    should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment - guidance 

RF emissions 

 

CISPR 11 

Group 1 

 

The SF-371 body fat scale    uses RF energy only for its internal 

function.  Therefore,  its  RF  emissions  are  very  low  and  are  not 

likely to cause any interference in nearby electronic equipment. 

 

RF emissions 

 

CISPR 11 

Class B   

 

The SF-371  body  fat  scale is suitable for use in all  

establishments,  including domestic establishments and those  

directly  connected to the public low-voltage power supply  netw

ork  

that supplies   buildings used for domestic  purposes. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Harmonic emissions 

 

 

IEC 61000-3-2 

N/A 

Voltage fluctuations / 

flicker emissions 

 

IEC 61000-3-3 

N/A 

 

Guidance and manufacturer's declaration – electromagnetic immunity –   
for all EQUIPMENT and SYSTEMS   

 

Summary of Contents for IVSCK

Page 1: ...batteries OPERATION 1 When in the first use please insert the batteries 2 Place the scale on a hard and flat surface Press the ON OFF button the scale will be zero in 3 seconds 3 Put the object on th...

Page 2: ...any chemical cleanser to clean the scale Clean your scale with a damp cloth but do not immerge your scale into any liquid as this can damage the inner parts 3 To ensure that the scales of life please...

Page 3: ...ase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radi...

Page 4: ...ment guidance 4 RF emissions CISPR 11 Group 1 The SF 371 body fat scale uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interferen...

Page 5: ...that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode N A Mains power quality should be that of a typical commercial or hospital environment...

Page 6: ...IEC 61000 4 8 NOTE UT is the a c mains voltage prior to application of the test level...

Page 7: ...N A 3 V m Portable and mobile RF communications equipment should be used no closer to any part of the SF 371 body fat scale including cables than the recommended separation distance calculated from th...

Page 8: ...the SF 371 body fat scale b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3V m Recommended separation distances between portable and mobile RF communications equipmen...

Page 9: ...equency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for t...

Reviews: