PRESSURE RELIEF ALGORITHM
•
Initial Phase of Regulation:
(when patient has yet to reach the
target temperature)
– 12 minutes of water flowing
– 1 minute of no water flow
•
Steady State:
(begins when patient reaches the
set point temperature)
– 12 minutes of water flowing
– 12 minutes of no water flow
•
At Any Point:
– Difference of >0.3°C between Set
Point and Core Temperature will
cause water to flow back into the
wrap