C LEAN ING , D IS INFE CT ION AND M AINT ENAN CE
ROUTINE MAINTAINENCE
Turn the ON/OFF switch to OFF.
1.
Clean and/or Disinfect Exterior:
A.
Battery Housing
The housing may be wiped down with water or a 10% bleach solution using a
soft cloth.
Note
: Avoid the use of acetone or other solvents that may damage the surface.
B.
Heater Unit
Wiping down the outside surfaces with a soft cloth moistened with water or a
10% bleach solution.
Note
: Avoid the use of acetone or other solvents that may damage the surface.
Open the Heater Unit and inspect the seal around the heater plates. If the seal is
intact, using a soft cloth wipe the heater plates with a 10% bleach solution and
dry. If the seal around the heater plates is
not
intact, do not use and return the
unit for service or replacement.
Note
: Do not use abrasive cleaners on the Heater Unit or heater plates.
2.
Visual Inspection:
Once the unit is cleaned, inspect all wiring and connections for fraying or worn
wiring. If noted, do not use; return the unit for service. If the wear is confined to
the Heater Unit, replace the heater.
WARNING
Do not submerge the Heater Unit.
Practice standard hospital policy when handling blood products. Treat
blood as if it were infected and clean up all spills immediately.
14