Table 11: Water Tank Overfilled- Draining the Water Tank
Observation
Action to be taken
Water tank
overfilled.
If it is necessary to drain the water tank because of overfilling, proceed as
follows:
1. Connect connecting tube to the right Quick Coupling Connector (under
the Core Sensor socket). The ThermoWrap
®
cannot be connected when
emptying.
2. Connect the special male connector to the connecting tube. (
See
3. Turn ON the Allon
®
device
4. Select Empty mode in Services and click Start.
5. Allow the excess water to drain into a receptacle, pail or sink.
6. When the desired water level has been reached, turn OFF the Allon
®
device.
Figure 36: ThermoWrap® Connecting Tubes and Special Male Connector