21
➢
Rapid
this method is suitable for samples
with a moisture content greater than 30%.
After starting, the temperature will exceed
the set value by about 30% for 2 minutes
and then settle at the set value. This is to
speed up the drying process.
Then confirm the selection with the key
.
2
– Warm up:
this function allows you to
preheat the heater before performing a drying
cycle.
➢
On,
preheating function enabled.
➢
Off,
preheating function disabled.
With the function enabled before executing the
drying cycle, the preheating screen will be
displayed. Press
to cancel the preheating
and start immediately with the drying cycle.
N.B. the active preheating function allows you
to obtain more repeatable results as each
cycle takes place with the same starting
condition of the heater.
3
– Startup mode:
this function allows you to
choose the mode for starting the drying
program.2
➢
Manual,
with this mode every preparation
operation for the start of the drying cycle
must be confirmed by the user that after
switching off the heater must press the
confirmation key.
➢
Automatic,
with this mode every
preparation operation for the start of the
drying cycle will be performed
automatically when the heater is switched
off.