Appendix
Operating manual NORA Alu
V2.13 E
Page 34 of 52
13 Appendix
13.1 Technical data
Dimensions and weights
- Length:
102.7 cm
- Width in closed condition
(internal dimension):
54.0 cm
- External width of the closed
carrier frame:
70.0 cm
- Width in open condition (internal
dimension):
89.7 cm
- External width of the spread
carrier frame:
104.4 cm
- External width at the handle:
73.0 cm
- Min. height:
116.8 cm
- Max. height:
178.1 cm
- Turning radius:
120 cm
- Weight without packaging:
approx. 51 kg
- Safe Working Load (SWL)
:
max. 200 kg (Nora Alu)
Electrical data
- Voltage supply of the lift:
24 Volt battery
- Max. energy consumption of the
lift:
10 A = 240 VA
- Applied part:
Type B
- Duration of treatment:
10%, 2 min permanent operation / 18 min pause
- Protection class
IPX 4
Ambient conditions
Operation
- Temperature range:
10 °C to 40 °C
- Relative humidity:
30% to 75%, non-condensing
- Atmospheric pressure:
800
–
1060 hPa
Storage and transport
- Temperature range (lift):
- 40 °C to 70 °C
- Temperature range (battery):
- 15 °C to 40 °C
- Relative humidity:
10% to 80%, non-condensing
- Atmospheric pressure:
500
–
1100 hPA