Beijing Choice Electronic Technology MD2000C User Manual Download Page 15

24

25

Fig.4.8

Inaccurate measurements may be caused by

• 

Significant levels of dysfunctional hemoglobin (such as carbonyl - 

hemoglobin or methemoglobin);

• 

Intravascular dyes such as indocyanine green or methylene blue;

• 

High ambient light. Shield the sensor area if necessary;

• 

Excessive patient movement;

• 

High-frequency electrosurgical interference and defibrillators;

• 

Venous pulsations;

• 

Placement of a sensor on an extremity with a blood pressure cuff, 

arterial catheter, or intravascular line;

• 

The  patient  has  hypotension,  severe  vasoconstriction,  severe 

anemia, or hypothermia;

• 

The patient is in cardiac arrest or is in shock;

• 

Fingernail polish or false fingernails;

• 

Weak pulse quality (low perfusion);

• 

Low hemoglobin;

Maintenance and Cleaning
WARNINGS!

• 

Disconnect the AC power before cleaning the monitor or sensor.

Note: 

1. Remember to save the settings or you can cancel the settings. 
2. Press default, all settings will be reset to factory defaults, please be 

carefully. 

Measurement principle

Principle  of  the  oximeter  is  as  follows:  A  mathematical  formula  is 

established  making  use  of  Lambert  Beer  Law  according  to  Spectrum 

Absorption  Characteristics  of  Reductive  hemoglobin  (RHb)  and 

Oxyhemoglobin  (HbO2)  in  glow  and  near-infrared  zones.  Operation 

principle  of  the  instrument:  Photoelectric  Oxyhemoglobin  Inspection 

Technology  is  adopted  in  accordance  with  Capacity  Pulse  Scanning 

and Recording Technology, so that two beams of different wavelength of 

lights (660nm glow and 905nm near infrared light) can be focused onto a 

human nail tip through a clamping finger-type sensor. A measured signal 

obtained by a photosensitive element, will be shown on the oximeter’s 

display through process in electronic circuits and microprocessor.
Measuring
a.Power on the device
b.Before  starting  a  measurement,  verify  that  you  have  selected  a 

suitable size sensor for the patient.
c.Connect the end of sensor to the device, the other end to the patient 

measured finger, as shown in Fig.4.7

Fig.4.7

Place  the  measured  finger  into  the  probe.  Several  minutes  later,  the 

result will display on the screen as shown in Fug.4.8

Summary of Contents for MD2000C

Page 1: ...Vital Sign Monitor Mode MD2000C Version 1 0 ...

Page 2: ...ny license under the patent rights of our company nor the rights of others Our company does not assume any liability arising out of any infringements of patents or other rights of third parties Our Company reserves the right to make changes in specifications and features shown herein or discontinue the product described at any time without notice or obligation Also content of the manual is subject...

Page 3: ...lobin or methemoglobin The presence of high ambient light Shield the sensor area with a surgical towel or direct sunlight for example if necessary Venous pulsations may cause erroneous low readings e g tricuspid value regurgitation Venous congestion may cause under reading of actual arterial oxygen saturation Therefore assure proper venous outflow from monitored site Sensor should be not below hea...

Page 4: ...ck the quantity and status of the battery Make sure the safety of the patient if it needs cut off the power supply take out the leads or sensor Do drop the fire goods medal or liquid into the equipment If these effective use of the monitor Notes contain important information that may be overlooked or missed Warnings Before use carefully read the manual Operation of the equipment may be affected by...

Page 5: ...ommunity hospitals and medical facilities 1 5 Electromagnetic Interference Under the normal measuring the equipment is not interference the surrounding people unit and environment During the process of sending data the device interference the surrounding people unit and environment If the equipment in the high frequency electromagnetic environment it will do harm to the equipment and the intended ...

Page 6: ...apter the light on 3 Indicator light turn on the device the light on 4 Power button press and hold this button for 3S to power on and 4S to turn off 5 Alarm silence button press the button to silence the audible alarm for one minute 6 Shortcut button of NIBP measuring 7 Menu button Press menu button enter into the setting menu 8 Rotary Knob The operator uses the rotary knob to select the menu item...

Page 7: ...o the charging port located on the back of the device then it becomes to charge the device When the battery reaches 100 the battery indicator become fullfilled Unplug the power adapter after finish charging The battery is new and fully charged the minimum operating time of the device is 3 5h The battery charge time from depletion to 90 charge is 2 5h The battery charge time from depletion to 100 c...

Page 8: ... confirm Patient information Fig 3 3 In this submenu fullfill in the patient information Fig 3 4 Note 1 Popup the keyboard when input the last name first name and ID CHAPTER 3 Setting First time use the device you should setup the following parameters after turn on the device Fig 3 1 Language English NIBP unit mmHg kPa Height unit cm in Weight unit kg Ib Date format yyy MM dd yyy MM dd dd MM yyy T...

Page 9: ... icon red cycle 1 then press to change case rotate the knob to the delete icon red cycle 2 then press to delete letter or number rotate the knob to the symbol icon red cycle 2 then press to the symbol keyboard 3 About the kind adult and Pediatric Alarm setting Before setting please enter password 2222 to set the parameter Fig 3 5 How do you input passwords 1 Rotate the knob to choose the item 2 Pr...

Page 10: ... color Fig 3 13 High limit SYS range is 16mmHg 295mmHg Low Limit SYS range is 15mmHg 294mmHg High limit DIA range is 11mmHg 285mmHg Low Limit DIA range is 10mmHg 284mmHg Unit mmHg kPa System setting Fig 3 10 Language English WIFI on off Bluetooth on off CO2 on off TEMP on off Height Unit cm in Weight Unit kg Ib About This Device Under the system setting rotate and press the knob on about to check ...

Page 11: ...ime setting screen Under the time setting screen rotate the konb on the time setup to adjust the date and time Rotate the knob on the delete that on the bottom of the screen press the knob to delete data Please take caution to the deletion of data you will never get the data back once deleted Reset to factory default Note All the setting will be reset to factory defaults Please be careful Fig 3 14...

Page 12: ...ply the cuff and its pressurization on the arm on the side of a mastectomy Pressurization of the cuff can temporarily cause loss of function of simultaneously used monitoring device on the same limb Check the operation of the Automated sphygmomanometer does not result in prolonged impairment of the circulation of the blood of the patient Any blood pressure reading can be affected by the measuremen...

Page 13: ...is not wrapped too tightly around the limb Excessive tightness may cause discoloration and eventual ischemia of the extremities The width of the cuff should be either 40 of the limb circumference or 2 3 of the upper arm length The inflatable part of the cuff should be long enough to encircle 50 80 of the limb The wrong size of cuff can cause erroneous readings If the cuff size is in question then ...

Page 14: ...activated when the continuous inflation last more than 30 seconds If there is no value when measurement time exceeds 120 seconds under the adult mode the measurement will be canceled You can press the START NIBP button to cancel a NIBP measurement when necessary Maintenance and Cleaning WARNINGS Do not squeeze the hose of cuff Do not allow liquid to enter the connector socket when cleaning the mon...

Page 15: ...e carefully Measurement principle Principle of the oximeter is as follows A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin RHb and Oxyhemoglobin HbO2 in glow and near infrared zones Operation principle of the instrument Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacit...

Page 16: ... Specifications Wavelength Radiant Power RED 660 3nm 3 2mw IR 905 10nm 2 4mw Note 1 The pulse oximeter equpment is calibrated to display functional oxygen saturation 2 Pulse oximeter monitor the pulse oximeter probe and probe cable extender have been validated and tested for compliance with this international standard 3 Please use the probe and cable that provided by our company 4 Please Verify th...

Page 17: ... to repair Repair for sponsoring and contributing machine customer should send the machine to us to repair Updating software is free 5 4 2 Exemption and restriction Warranty does not apply to the damage or loss sustained due to well known act of god such as fire earthquake flood thunder cyclone hail electrical storm blast building collapse commotion CHAPTER 5 MAINTENANCE 5 1 Examination Before use...

Page 18: ...g the instrument or parts b Damage or loss sustained due to inspected or repaired by other institute that is not certified c Damage or alteration by anyone other than our company authorized service personnel The damage or lose sustained due to connection to peripheral equipment such as printer computer etc that are not provided by our company are not covered by the warranty Obligation restriction ...

Page 19: ...authorized by our company The illustrations used in this manual may differ slightly from the appearance of the actual device The maximum application time for each type of probe at a single site is 4h The equipment has been calibrated users do not to calibrate In order to ensure the accuracy of the probe please change the probe once a year Make sure that the type of probe need to be specified The e...

Page 20: ... applied part According to the degree of protection against dust and water IP22 According to the mode of operation continuous operation Applied part SpO2 probe Temperature probe NIBP cuff Note All the applied part contains temperature probe cuff electrode plate and SpO2 probe are compliance with the biological compatibility Standard List of accessories Accessories Mode Type Quantity SpO2 probe M 5...

Page 21: ...onsult the dealer or an experienced radio TV technician for help CHAPTER 7 Compliance Information for EMC Test Electromagnetic emission Emission test Compliance Electromagnetic Environment guidance Conducted and radiated RF emissions Group 11 Group 1 Class B Home Healthcare environment Electromagnetic immunity Immunity test IEC 60601 test level Compliance Level Electromagnetic Environment guidance...

Page 22: ...1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3 28 1845 1970 2450 2400 2570 Bluetooth WALN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217Hz 2 0 3 28 5240 5100 5800 WLAN 802 11 a n Pulse modulation b 217Hz 2 0 3 9 5500 5785 Manufactured for Beijing Choice Electronic Technology Co Ltd Room 4104 No A12 Yuquan Road Haidian District 100143 Beijing PEOPLE S REPUBLIC OF...

Reviews: