33
Calibration
successful
Calibration failed
Retry calibration
Firmware version
N/A
= Not Available on this device
Troubleshooting
The unit fails to turn on
If the unit fails to turn on properly, check if
the low or empty battery symbols are
shown or replace the batteries. Ensure that
the batteries are inserted the correct way
around, matching the symbols moulded
into the plastic.
High readings recorded for reported non-smoker
If a breath test on a non-smoker gives an indication of CO being present, it may indicate the following:
1.
A high ambient level of CO (possibly due to a faulty combustion appliance or exhaust fumes).
2.
The effects of passive smoking.
3.
The device is contaminated by alcohol or another organic substance.
4.
The patient may have a very high level of hydrogen on their breath, which can be generated in the
human digestive system.
5.
The sensor may have drifted out of specification (see below).