35
Data Reset
To access the Data Reset button, first remove the
back cover and the breath drying cartridge.
This will reveal the Data Reset button at the top
right of the NObreath®.
Press and hold the Data Reset button for 5
seconds and the screen will reveal the reset icon.
NOTE:
A data reset will erase all patient data from
the device the PIN will be reset to the default,
0000.
Press the tick to confirm the data reset or cross to
cancel.