background image

 

 

30 

 

 

Warranty 

Bedfont® Scientific Limited warrants the Micro

+

™ Smokerlyzer® produ

ct (excluding batteries) to be 

free of defects in materials and workmanship for a period of five years from the date of shipment. 

Bedfont’s sole obligation under this warranty is limited to repairing or replacing, at its choice, any 

item covered under this warranty when such an item is returned, intact and prepaid, to Bedfont® or 
the local representative. 

These warranties are automatically invalidated if the products are repaired; altered, void 
labels removed or otherwise tampered with by unauthorised personnel, or have been 
subject to misuse, neglect or accident. 
 

Never dispose of any electronic instrument or batteries in domestic waste. At the end of the 

product’s life, contact Bedfont® or its distributor for disposal instructions.

 

 

Returns 

If the equipment requires servicing, please contact the local Bedfont® customer service department, 
distributor or supplier before returning any goods. 

1.

 

Supply the customer repairs department with the monitor serial number and description 
of the fault. A returns/ ticket number will then be assigned. Please state the 
returns/ticket number on the outside of the box when returning the monitor, ensure 
that a telephone number, fax number, and full return address are clearly stated. 

2.

 

The product must also be decontaminated before it is returned according to the local 
regulations. Bedfont® can provide a decontamination certificate to complete, which also 
needs to be attached to the outside of the box. Failure to do so will result in the product 
being subject to Bedfont® decontamination procedure and will delay the service/repair. 

3.

 

Bedfont® advises that a courier service is used when returning monitors. This enables 
insurance for goods against loss or damage in transit. When the goods are received, an 
email will be sent stating so.  

4.

 

If the device has been returned for repair it will then be examined and 

an engineer’s 

report will be sent with a quotation for the repair, which will include an authorisation 
form. Complete the authorisation form, and ensure that 

the ‘official purchase order 

number’

 is included. Please contact the customer repairs department if  an 

‘official 

purchase order number’

 cannot be supplied. 

5.

 

If the  

monitor is still in warranty and the fault is covered by warranty, see ‘warranty’ 

section 

of this manual, Bedfont® will repair it and return it with an ‘engineer’s report’, 

free of charge.  

6.

 

If the decision is made to not to proceed with the repair, a handling fee will be charged. 
Ensure the completed authorisation form is returned 

with an ‘offi

cial purchase order 

number’.

 

Summary of Contents for Micro+ Smokerlyzer Series

Page 1: ...Smokerlyzer Range User manual Our family innovating health for yours For use with Micro Smokerlyzer ...

Page 2: ...eld upright at all times during a breath test WARNING Do not block the exhaust ports on the device at any time Blocking the exhaust ports may cause erroneous readings WARNING Avoid exposure to known sources of EMI electromagnetic interference such as Magnetic Resonance Imaging MRI systems diathermy lithotripsy electrocautery RFID Radio Frequency Identification and electromagnetic security systems ...

Page 3: ... 8 Patient Profiles 9 Reviewing history 11 Changing date and time 11 Changing breath holding time 12 Changing breath testing thresholds 14 Maintenance Reminders 15 Maintenance 16 Calibration 17 Technical Specification 21 Safety Information and Device Symbols 22 Environment 22 COdata software connecting to the PC 23 Buttons 24 Troubleshooting 27 Warranty 30 Returns 30 Responsible Manufacturer and C...

Page 4: ...which if not avoided may result minor or moderate injury CAUTION indicates a potentially hazardous situation which if not avoided may result in damage to the device NOTE used to call attention to notable information that should be followed during use Compliance The Micro Smokerlyzer monitor is CE marked according to the Medical Device Directive 93 42 EEC as amended by 2007 47 EC Please refer to th...

Page 5: ...nt Layout 1 Power button 2 USB port for use with COdata software 3 Display 4 D piece aperture 5 Battery compartment clip 6 Battery compartment 7 Exhaust port 8 Manufacturer label 9 Exhaust port 1 2 4 3 5 6 7 8 9 ...

Page 6: ...eath test 4 Ambient air test 5 Patient profiles 6 Settings Taking a breath test Attach a breath sampling D piece and new SteriBreath Eco mouthpiece Turn on the monitor by pressing the power button once Press the breath test icon on screen either the male or female symbol 1 2 3 4 5 6 ...

Page 7: ...6 Inhale and hold breath for the pre set 15 second countdown Press the home button at any time to cancel the breath test A beep will sound during the last three seconds of the countdown ...

Page 8: ...o mouthpiece aiming to empty lungs completely The ppm and equivalent COHb and or FCOHb levels will rise and hold onscreen When the test is finished the volume home and save icons will appear at the bottom of the screen ...

Page 9: ...er button for 3 seconds unit will also power off after 2 minutes of inactivity to save power Conducting an ambient test The environmental air can be tested to check for CO in the environment using the Micro Smokerlyzer This is particularly useful in the event of a patient reporting a non smoker status but providing a breath reading of that of a smoker suggesting they have inhaled environmental CO ...

Page 10: ...an lead to false negative readings Patient Profiles All patient profiles saved on the device will be listed here To view patient information press the relevant patient profile block If start breath test is selected from this screen the breath test will be automatically saved to the profile Once the type of breath test has been assigned to the patient only one symbol will appear ...

Page 11: ... Click the save icon to confirm or the home icon to return to the home screen To access the patient s breath test results history press the graph icon This will then bring up the patient s testing history To delete a patient profile press the delete icon To confirm press the delete icon again ...

Page 12: ... to access this press the settings icon Bedfont does not recommend that the environmental test alone is used as a measure of passive smoking as the very high dispersiveness of CO at the levels involved with smoking can lead to false negative readings Changing date and time Press the settings icon ...

Page 13: ...mat and 12h or 24h for the time format Dates and times are then adjusted by selecting the number needing changed and pressing the up or down arrow icons to increase and decrease Press the save icon to save the new settings Changing breath holding time Press the settings icon ...

Page 14: ...13 Press the breath holding time icon Use the arrows to adjust the breath holding time Press the save icon to save changes ...

Page 15: ...Borderline 7 9 1 Red Smoker low addicted 10 15 2 Red Smoker moderately addicted 16 25 3 Red Smoker heavily addicted 26 35 3 Red Flashing Smoker very heavy addicted 36 This table shows the pre set breath testing thresholds Press the settings icon To change the breath test thresholds press the traffic light icon ...

Page 16: ...reminder The D piece requires changing every 30 days This reminder is displayed at switch on Sensor change reminder This reminder is displayed at switch on when the sensor is nearing its change date After 60 days of this reminder being displayed the red date alarm takes over ...

Page 17: ...t all settings to the factory defaults After performing a reset the device will need to have the date time set and be calibrated before it can be used 7 Replace batteries when indicated by the empty symbol 8 Bedfont recommend removal of the batteries when the device is not used for prolonged periods of time to prevent leakage 9 Replace breath sampling D piece every 30 days or if visibly soiled or ...

Page 18: ...ning agents containing alcohol or other organic solvents as long term exposure to these vapours will damage the CO sensor inside 3 Under no circumstances should the instrument be immersed in liquid or splashed with liquid Calibration The Smokerlyzer monitors must be calibrated within 17 25 C If it is too cold to calibrate 17 C a blue thermometer will be shown onscreen Move the Smokerlyzer to a war...

Page 19: ...able Once this has been successfully carried out the device will perform a zero do not apply gas at this stage The first step of the calibration process will be shown onscreen follow the onscreen instructions to finish the calibration process ...

Page 20: ...rew the fine control valve and flow indicator assembly to the gas can This is best done by screwing the gas can into the valve 3 Press breath test symbol on screen to start a breath test 4 When prompted by the exhale symbol open the fine control valve or regulator ...

Page 21: ... monitoring the rate of flow 7 If the final displayed value is less than 18ppm or higher than 22ppm stop the test and perform a calibration or send back to Bedfont for a free of charge calibration following the instructions A successful calibration will be indicated by the tick icon press the home icon to return to the home screen ...

Page 22: ...r 3 x AA LR6 or equivalent up to 1000 minutes 1 x CR2032 Lithium coin cell T90 response time 30 seconds Operating temperature 15 40 C Storage transport temperature 0 50 C Operating storage transport pressure Atmospheric 10 Operating humidity 15 90 RH non condensing Storage transport humidity 0 95 non condensing Sensor operating life 5 years Sensor sensitivity 1ppm Sensor drift 5 per annum Dimensio...

Page 23: ...spose of according to WEEE Serial Number Unique device identifier Indicator of Medical Devices Magnetic Resonance MR unsafe Manufactured by and date Bedfont Logo Environment The Micro Smokerlyzer product complies with the Medical Device Directive 93 42 EEC as amended by 2007 47 EC and the Medical Device Electromagnetic Compatibility standard EN 60601 1 2 WARNING Portable RF communications equipmen...

Page 24: ...Hz 2 7GHz 385 MHz 27 V m PM 18 Hz 450 MHz 28 V m FM 1 kHz sine 710 MHz 9 V m PM 217 Hz 745 MHz 9 V m PM 217 Hz 780 MHz 9 V m PM 217 Hz 810 MHz 28 V m PM 18 Hz 870 MHz 28 V m PM 18 Hz 930 MHz 28 V m PM 18 Hz 1720MHz 28 V m PM 217 Hz 1845 MHz 28 V m PM 217 Hz 1970 MHz 28 V m PM 217 Hz 2450 MHz 28 V m PM 217 Hz 5240 MHz 9 V m PM 217 Hz 5500 MHz 9 V m PM 217 Hz 5785 MHz 9 V m PM 217 Hz Met or exceeded...

Page 25: ...ted to the PC and switched on Double click the COdata icon on the PC to start the programme Refer to the supplied documentation for how to operate COdata Buttons Description Micro Battery condition full Battery condition low Battery condition empty Breath test Environmental test Settings Patient profiles View patient results ...

Page 26: ... Inhale Hold breath Countdown timer Exhale ppm reading COHb reading Show FCOHb reading FCOHb reading Ambient CO reading Home Change D piece Change traffic light cut off levels Save Change breath hold time ...

Page 27: ...re Countdown to sensor change Change sensor Calibrate device Attach flow mete to gas canister Attach calibration adaptor to D piece Attached D piece to monitor and turn on gas flow Unit calibrating Calibration successful Calibration failed Retry calibration ...

Page 28: ... non smoker If a breath test on a non smoker gives an indication of CO being present it may indicate the following 1 A high ambient level of CO possibly due to a faulty combustion appliance or exhaust fumes 2 The effects of passive smoking 3 The device is contaminated by alcohol or another organic substance 4 The patient may have a very high level of hydrogen on their breath which can be generated...

Page 29: ...nt check gas or send back to Bedfont The check gas required is Bedfont 20ppm carbon monoxide in air procedure below 1 Ensure the fine control valve is in the off position 2 Screw the fine control valve and flow indicator assembly to the gas can This is best done by screwing the gas can into the valve 3 Screw the fine control valve and flow indicator assembly to the gas can This is best done by scr...

Page 30: ... per minute 7 Allow the gas to flow through the instrument for the duration of the test again monitoring the rate of flow 8 If the final displayed value is less than 18ppm or higher than 22ppm stop the test and perform a calibration or send back to Bedfont for a free of charge calibration following the instructions ...

Page 31: ...onitor ensure that a telephone number fax number and full return address are clearly stated 2 The product must also be decontaminated before it is returned according to the local regulations Bedfont can provide a decontamination certificate to complete which also needs to be attached to the outside of the box Failure to do so will result in the product being subject to Bedfont decontamination proc...

Page 32: ...elevant paperwork A carriage fee will be charged if the monitor is no longer in warranty Responsible Manufacturer and Contacts Bedfont Scientific Ltd Station Yard Station Road Harrietsham Maidstone Kent ME17 1JA United Kingdom www bedfont com ask bedfont com 0044 1622 851122 ...

Page 33: ...change or update his literature without prior notice Registered in England and Wales Registered No 1289798 Stephen Rowe Cristimar E4 1 Ave Juan Carlos I Los Cristianos Arona 38650 Santa Cruz de Tenerife Spain Bedfont Scientific Ltd Station Road Harrietsham Maidstone Kent ME17 1JA England Tel 44 0 1622 851122 Fax 44 0 1622 854860 Email ask bedfont com Web www bedfont com Bedfont Scientific Limited ...

Reviews: