Open the Bag Clamp and press the next arrow to
continue.
An hourglass will display onscreen as the device
will automatically calibrate, drawing the gas
from the CaliBag®.
A successful calibration is shown by a green tick.
Press Save
to complete the calibration.
An unsuccessful calibration is shown by a red cross.
If an unsuccessful calibration is shown by a red
cross, contact MGC Diagnostics Corporation or the
local distributor.
Ensure both Bag Clamps are closed before
removing the CaliBag® and remove all adapters
from the FeNObreath™️.
Remove the Bung from the FeNObreath™️.
11
10
9
12
13
14
Please ensure the FeNObreath™ monitor is
switched off before removing the back cover.
Using the screwdriver provided with the
FeNObreath™, unscrew the screw on the back of
the monitor.
Remove the back cover by sliding
and lifting it off.
To remove the scrubber, first press down on
clip at the top left-hand side of the monitor.
Slide out the scrubber then lift to remove.
Remove all packaging from the new NO
Scrubber and insert it into the FeNObreath™️.
Re-attach the back cover of the
FeNObreath™️ monitor.
To secure the FeNObreath™ back cover, use the
screwdriver to resecure the screw into place.
How to replace the NO Scrubber
4
5
6
7
1
2
3
8
(
CAUTION:
NO scrubber contains potassium permanganate and
should not be tampered with and should not be exposed to
skin.)