D. Calibration
1. Ensure you are performing a calibration in a fresh air environment.
2. Turn piCO+
™
ON
by holding the button down for three seconds.
3. Main Screen
a.
Single click
the button to move the
pointer to
the
calibration canister icon.
b. Next,
double click
the button to select the
calibration canister icon.
4. Calibration Mode
a.
Double click
button to select the
canister icon
.
b.
Immediately
turn on the gas by turning the valve.
No fine tuning
is required because this regulator is fixed at
1 lpm
.
c. T
he monitor’s reading will slowly climb from zero and
an hour glass will be shown on the screen as well.
d. When you receive a
check mark, turn
OFF
the gas.
e.
Double click
the button to select
return arrow and to go previous
menu.
f.
Double click
the button to select the
return arrow and return to
main menu.
g.
Calibration Complete
Note: If you receive and
instead of a
please contact coVita for
further assistance
.