Two-Month Maintenance
Replace CTS Blade/Wick (CTS Option only)
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-40
October 2005
5
Refer to
Remove the blade clamp, blade (1), and wick (3) as follows:
• Hold the blade (1) at the top of its shaft. Do not touch the points (2) of
the blade.
• Lift up the blade and remove it with the wick at the same time:
- Move the blade a small distance laterally, if necessary, to release the
wick from its socket.
- Do not touch the wick because it contains CTS Auto-Gloss lubricant
which is very slippery and difficult to clean.
- The wick (3) should be attached to the blade.
• Discard the old blade clamp, blade, and wick into a biohazard, sharps
container.
Figure 10.23
6
Refer to
Note: A new blade is shipped with a wick attached. The points of the blade
are very sharp. The blade shaft is not sharp.
Remove a new blade and wick from its container.
• Hold it at the top of the blade shaft, immediately under the blade clamp
as shown in
Step
Action
, continued
(3 of 5)
A007368P.EPS
3
1
2