Reagent Load
Load/Unload Reagent
Reagent Load/Calibration
UniCel DxC Systems Instructions For Use A13914
Page 5-6
October 2005
Figure 5.5
DxC 600/800 Center Compartment
Modular chemistry reagent bottles are selected from the Rgts/Cal screen and replaced
as necessary according to volume. Reagent volume levels are automatically tracked by
the system after the initial volume is set. When a new reagent bottle is replaced the
system assumes the bottle is 100% full. If the new bottle is not 100% full, the volume
may be adjusted.
Load/Unload Modular Chemistries (MC)
1. Wash Concentrate II
2. No Foam solution
3. CTS Auto-Gloss
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select the Modular Chemistries <MC Rgts>.
3
Select Load [F1].
4
Open the leftmost and/or center doors to access modular reagents. Refer to
, and
(1 of 3)
A015922P.EPS
1
3
2