System Status and Commands
Shutdown
10
6
.
ON/OFF Switch
The circuit breaker/power ON/OFF (O/|) switch is located behind the front right-hand door of the
instrument.
Figure 10.1
ON/OFF (O/|) Switch
Line Converter Device
All power to the instrument is passed through the line converter device located in the lower right
compartment of the instrument. The line converter serves as a conditioner to limit the effect of line
voltage changes that could damage the sensitive electronic components of the system.
DC Output Power Supplies
Power is routed to redundant DC output power supplies which provide 24V or 36V power for
distribution throughout the system. By using redundant power supplies, not all are used at any one
particular time. Should one fail, the operator is alerted, but may continue operation unaffected.
Uninterruptible Power Supply (UPS)
The computer console requires an additional power source. The UPS plugs into a standard 110-120V
AC, 50/60 Hz power source. The UPS in turn supplies power to the computer console and the
1.
Main Power ON/OFF (O/|)Switch
A016487P.EPS
1
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......