A13914AF
9-125
Maintenance
As-Needed/As-Required Maintenance
9
10
Place four sample tubes that are capped and empty into an appropriate sample rack.
•
If the system is a DxC 600/800, go to Step 11.
•
If the system is a DxC 600i, go to Step 12.
11
For a DxC 600/800 system, place the rack into the autoloader with no sample programming.
Press the
RUN
button to pierce the tube caps. Inspect the bottom of each cap to confirm that the
cap was pierced.
12
For a DxC 600i system, loosen the two Phillips screws on the cover of the DxC Load Area and
remove the cover. Place the rack into the DxC Load Area with no sample programming.
NOTE
Do not load the rack into the DxC 600i Load Area.
Loosen the screw and open the cover of the DxC Load Area
RUN
button (located to the right of
the DxC 600i Load Area). Press the DxC Load Area
RUN
button to pierce the tube caps. Inspect
the bottom of each cap to confirm that the cap was pierced. Close the DxC Load Area
RUN
button
cover and tighten the screw. Replace the DxC Load Area cover.
Replace ISE Drain Pump Tube
The DxC system must be in
Standby
.
Materials Required:
•
ISE Drain Pump Tube (PN A43830)
1
Select:
•
Utils
from the menu bar
•
2 Maintenance
•
5 Hydropneumatic Maintenance
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......