9-112
Maintenance
As-Needed/As-Required Maintenance
Wipe the touchscreen with 70% isopropanol, water, or any commercial computer screen
cleaner.
Decontaminate the Sample Racks
1
Prepare a 10% bleach
*
solution (one part 5.25% sodium hypochlorite (PN A32319), nine parts
deionized water). Use within 24 hrs.
2
Using a large container or sink, immerse all sample racks in the bleach solution.
3
Allow the sample racks to soak at room temperature in the 10% bleach solution for 15–20
minutes.
4
Remove the racks, rinse with deionized water or tap water, and allow to dry.
IMPORTANT
Repeated decontamination of racks may result in rack label damage. If the label starts to
bubble or peel, remove and replace with a new label.
Decontaminate the Instrument
This procedure provides instructions for cleaning and decontaminating system tubing, waste
sumps, and reservoirs. Perform this procedure when microbial contamination is either evident or
suspected. This procedure should take about 1.5–2.0 hours.
CAUTION
All biohazard precautions should be observed when doing maintenance, service,
or troubleshooting on the system. This includes but may not be limited to wearing
gloves and eye shields, and washing hands after working on contaminated
portions of the system.
Materials Required:
•
10% bleach solution (one part 5.25% sodium hypochlorite (PN A32319), combined with nine
parts deionized water) - 8 Liters
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......