A13914AF
9-66
Maintenance
As-Needed/As-Required Maintenance
3
Locate the Chloride electrode (1). Remove electrode by turning the electrode retaining nut.
Figure 9.44
4
Remove the retainer nut from the old electrode. Remove the quad-ring (1) from the tip of the
old electrode. Inspect the electrode port if the quad-ring is not on the electrode. Discard the
quad-ring.
Figure 9.45
5
Unscrew the old tip from the electrode assembly. Discard the O-ring and the old electrode tip.
6
Install a new O-ring on chloride electrode tip and screw the tip onto electrode body.
7
Completely dry the electrode port with lint-free tissue.
A015946P.EPS