Maintenance
Monthly Maintenance
Figure 9.16
11
Apply a thin coating of Silicone Compound (PN 879049) over the tip of the electrode. Carefully
wipe the electrode end parallel to the gap with lintless tissue to remove all evidence of
compound. Rub to a bright finish.
IMPORTANT
Thoroughly wipe off the electrode surface to remove any residual silicone compound.
Excess compound could prevent a successful calibration.
12
Using lintless tissue saturated with deionized water; clean the electrode mounting port.
13
Using dry, lintless tissue, wipe the electrode mounting port.
14
Reinstall the quad-ring on the face of the electrode.
15
Reinstall the retainer and retainer nut on the BUNm/UREAm electrode.
16
Rotate the electrode until the electrode key enters the retainer keyway and reinstall the
retainer nut on the electrode-retainer assembly.
17
Align the keyway on the body of the retainer containing the electrode with the key in the
electrode port and install electrode assembly. Finger-tighten retainer nut.
18
Reconnect the pin lead into the connector panel by aligning the latch on the connector with the
notch on connector panel. Push the pin lead into the connector panel until the latch clicks.
Refer to
.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......