A13914AF
9-22
Maintenance
Monthly Maintenance
•
Type
5
in the
Number of primes to repeat
field.
•
Select
Start Prime
. (Prime cycles are completed when
Standby
appears in the system status.)
4
Observe the fluid level during prime cycles. As the damper becomes half full, turn the damper
to the upright position.
5
Wipe dry any reagent spill on the damper assembly.
6
Clip the damper assembly back into the upright position.
CAUTION
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
7
Lower the ISE module by releasing the locking pin. Be careful NOT to pinch tubing as the module
is lowered.
8
Reinstall the ISE module cover.
9
If the fluid level is above the MAX line:
Remove input line #30 (1) feeding into the damper.
•
Using a clean transfer pipette, remove excess reagent until damper is one-half full. Discard
the removed fluid.
•
Reconnect input line #30 to the damper, wipe dry any reagent spill on the damper top.
Figure 9.11