Troubleshooting
Cytometer Messages Table
13
Change Font
To change the font:
Format > Font
, and select the new font.
Find Help Topics
To find Help on other topics:
Help > Help Topics
.
Cytometerarchive.log File
T
he cytometerarchive.log file contains messages that were automatically moved from the
cytometer.log file after 30 days. This log is for Service use.
Cytometer Messages Table
lists the messages in alphabetical order, with their cause and what to do about them.
These are the messages produced by the Cytometer.
contact your local Beckman Coulter Representative if:
•
The recommended action does not solve the problem.
•
You need help.
Table 13.2
Cytometer Messages
Message
Probable Cause
Recommended Action
Acquisition Yield Warning
The threshold events exceeds the
processed and displayed events.
1.
Dilute the sample .
2.
Change the discriminator setting.
3.
Change the sample flow rate to medium
or low.
4.
Restart the entire system. If error
persists, contact your local Beckman
Coulter Representative.
Ambient Sensor Error
The ambient temperature sensor
failed during acquisition and the
acquisition was stopped.
If error persists, contact your local Beckman
Coulter Representative.
Ambient Sensor Warning
There has been no change in the
measured ambient temperature for
some time and the sensor may have
failed.
If warning persists, contact your local
Beckman Coulter Representative.
Ambient Temperature Warning
Ambient Temperature is outside the
system's specifications.
Reduce or wait till ambient temperature is
within specification.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......