Using Navios Software
Analysis Menu
Regions Introduction
A Region identifies a specific population of cells.
Up to 256 regions (total) can be set on any combination of Parameters, 32 of these are usable as
Gates.
Regions are drawn on a Single Parameter Histogram by setting the position of an upper and lower
boundary (Region markers).
Regions are drawn on a 2-Dimensional Dot Plot by enclosing a subpopulation of events with
rectangles, polygons (nonrectangular Regions) or using quadrant markers (four rectangular
Regions, set with only one point).
To speed up the analysis of data, when a Single Parameter Histogram Region or a Rectangular or
Polygonal Region is created, it is also assigned to a Gate. This assignment can be accepted or edited
later. The automatic assignment of Regions to Gate can be switched on or off if desired, using
Gates >
Region Properties
on a region to make it active and then click the right mouse button on it to display the Region
Properties dialog box.
.
The
Re-Calculate AutoGate
option allows a return to AutoGating after manual manipulation of an
Auto region. All the AutoGates are re-calculated, including positives analysis in a protocol. Autogate
functionality, including Positive Analysis, is suspended whenever the gate or positive analysis
region is edited during acquisition.
Polygonal Region
The Polygonal Region allows you to draw a free-form Region with up to 100 points and can include
horizontal acute angles on a dual parameter plot. To draw a Polygonal Region in a dual parameter
display see,
.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......