Beckman Coulter M502-22 Instructions For Use Manual Download Page 35

APPENDIX B

References

References

1. CLSI. Protection of Laboratory Workers from Occupationally Acquired Infections;

Approved Guideline-Second Edition. CLSI document M29-A2 [ISBN 1-56238-453-8].
CLSI, 940 West Valley Rd, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2001.

2. CDC. Recommendations for Prevention of HIV Transmission in Health Care Settings.

MMWR (Suppl. No. 2S):2S-18S, 1987.

3. CDC. Updated: US Public Health Service Guidelines for the Management of Occupational

Exposures to HBV, HCV and HIV and Recommendations for Post Exposure Prophylaxis.
Appendix A and B. MMWR 50 (RR-11): 1-42, June 29, 2001.

55-005187-001EC

B-1

Summary of Contents for M502-22

Page 1: ...Use StatSpin Express 3 Centrifuge Model Number M502 22 This manual is intended for SSX3 Express 3 for 100 to 240 VAC 50 60 Hz 55 005187 001EC September 2018 Beckman Coulter Inc 250 S Kraemer Blvd Bre...

Page 2: ...emarks Beckman Coulter the stylized logo and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter Inc in the United States and othe...

Page 3: ...ations Manual to a Beckman Coulter Instructions for Use IFU manual and made general clarifications to the IFU Added Revision History Safety Notice Symbols and Definitions table Alerts for Warning Caut...

Page 4: ...Revision History iv 55 005187 001EC...

Page 5: ...not avoided could cause death or serious injury Warning can indicate the possibility of erroneous data that could cause an incorrect diagnosis Caution Caution indicates a potentially hazardous situat...

Page 6: ...materials to damage See Specifications Warning Do not expose the rotor to strong or concentrated acids bases esters aromatic or halogenated hydrocarbons ketones or strong oxidizing agents or environm...

Page 7: ...between the system and other equipment Beckman Coulter recommends the following actions to correct the interference This equipment complies with the emission and immunity requirements described in th...

Page 8: ...eferences Caution Use the same size and style tubes in opposite positions Balance liquid in tubes to within 0 5mL If glass tubes are used to balance rotor load replace the glass balance tube every ten...

Page 9: ...cient vigor to cause a hazard Caution This product may contain a chemical known to the State of California to cause cancer or birth defects or other reproductive harm Caution Ensure to replace the fus...

Page 10: ...to consult the instructions for use ISO 15223 1 Medical devices Symbols to be used with medical device labels labelling and information to be supplied Part 1 General Requirements 5 4 3 Caution Indica...

Page 11: ...t 13 2005 2 The device is not to be disposed of via the municipal waste collection system of any member state of the European Union Customers must understand and follow all laws regarding the correct...

Page 12: ...06 Marking for Control of Pollution Caused by Electronic Information Products requirements OFF power To indicate disconnection from the mains at least for mains switches or their positions and all tho...

Page 13: ...en the medical device was manufactured ISO 15223 1 Medical devices Symbols to be used with medical device labels labelling and information to be supplied Part 1 General Requirements 5 1 3 Catalogue Nu...

Page 14: ...of two minutes at 7200 rpm 3 Minute Cycle The 3 min button selects a cycle of three minutes at 7200 rpm 5 Minute Cycle The 5 min button selects a cycle of five minutes at 5600 rpm Error or Service in...

Page 15: ...nd Closing the Cover 3 1 Cover Interlock By pass 3 1 Removing and Installing the Rotor 3 1 Removing the Rotor 3 1 Reinstalling the Rotor 3 2 Spinning the Sample 3 2 Loading 3 2 Cycle Selection 3 3 Tub...

Page 16: ...Contents xvi 55 005187 001EC...

Page 17: ...One Instructions for Use Manual One manual latch release tool One Setting Guide One 5 16 inch hex key tool Installing the System 1 Place the StatSpin Express 3 on a level surface suitable for laborat...

Page 18: ...e the two minute cycle light on the front keypad Press the stop button to release the latch and open the cover Warning Outside of North America do not use the power cord supplied Use power cord for at...

Page 19: ...and 10 mL which require lower g forces Five minutes at 5600 rpm produces the same degree of separation achieved with three minutes at 7200 rpm Operator Controls Express 3 Operator Controls Table 2 1...

Page 20: ...ed indicator flashing Rotor failed to reach full speed in 30 seconds Confirm no obstruction present in bowl Contact customer service if error continues to occur 3 min light and red indicator flashing...

Page 21: ...rotor and centrifuge is on level surface Remove and inspect rotor for cracks or warping Contact customer service if error continues to occur 3 min and 5 min lights and red indicator flashing Locked ro...

Page 22: ...erts for 10 mL tubes 16 x 100 mm 8 pk Included with centrifuge SV08 Inserts for 5 mL tubes 13 x 75 mm 8 pk Included with centrifuge SV09 Inserts for 7 mL tubes 13 x 100 mm 8 pk SV10 Insert for Greiner...

Page 23: ...ion The cover interlock bypass is for emergency use only Disconnect the power cord from the wall outlet and ensure the rotor has come to a complete stop before using the interlock bypass If the equipm...

Page 24: ...e the rotor correctly can damage the centrifuge Loading 1 Be sure the correct tube inserts are installed The StatSpin Express 3 centrifuge is provided with eight tube inserts for 13 x 75 mm tubes Prod...

Page 25: ...n equivalent to 3 min setting Centrifugal forces are decreased to accommodate larger glass tubes which require lower g forces 5600 rpm 2685 xg Tube Guide by Settings Tube Guide by Settings Table 3 2 T...

Page 26: ...Operating Instructions Spinning the Sample 3 4 55 005187 001EC...

Page 27: ...iner should be replaced monthly or whenever a spill occurs Clean the outside surfaces and control panel with a water dampened cloth and mild detergent Clean the inner surface or bowl with a mild deter...

Page 28: ...h solution 6 Do not attempt to clean an insert that has a broken tube Dispose of the insert immediately 7 Dry the rotor and inserts with a clean absorbent paper towel or allow to air dry 8 Inspect rot...

Page 29: ...d below the ON and OFF switch 3 Replace the fuse 5 x 20 mm type rated F3 0A 250V 4 Reinstall the fuse block and reconnect the power cord Troubleshooting No Power Light Table 4 1 No Power Light Is the...

Page 30: ...oulter Customer Support at 1 800 854 3633 Be sure to complete and return the warranty card as directed Decontamination before returning for service Any instrument or accessory containing accumulated b...

Page 31: ...ves salespersons or agents of Beckman Coulter which are inconsistent or in conflict with or in addition to the terms of this Agreement will not be binding upon Beckman Coulter unless reduced to writin...

Page 32: ...Maintenance Service 4 6 55 005187 001EC...

Page 33: ...th 10 in Weight 8 3 4 lbs Environmental Indoor use Operating temperature 2 C to 40 C Maximum 80 RH between 2 C to 30 C decreasing to 50 RH at 40 C Maximum altitude up to 2 000 m Main supply voltage fl...

Page 34: ...Specifications Specifications A 2 55 005187 001EC...

Page 35: ...alley Rd Suite 1400 Wayne Pennsylvania 19087 1898 USA 2001 2 CDC Recommendations for Prevention of HIV Transmission in Health Care Settings MMWR Suppl No 2S 2S 18S 1987 3 CDC Updated US Public Health...

Page 36: ...References References B 2 55 005187 001EC...

Page 37: ......

Page 38: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...

Reviews: