PN B05577AC
6-59
Sample Modules
Tornado Dry Powder System
6
Cleaning the Cell Windows
To clean the windows the sample cell must be removed from the module housing.
1
Loosen the screws that hold the bracket in place. Lift the bracket to disengage the pins from the
pin slots. See
.
Figure 6.36
Bracket Screws
2
Undo the four (top and bottom) latches found in the back of the module that attach the sample
cell to the module housing (
1.
Pin
2.
Bracket
3.
Slot
4.
Screw Securing the Bracket