PN B05577AC
6-17
Sample Modules
Aqueous Liquid Module
6
Replacing the Diffraction Cell Window
Equipment Needed
•
lens tissues
•
lens cleaning solution
•
window sealing tool
Procedure
Use this procedure to replace a diffraction sample cell window that is scratched or damaged.
To replace a diffraction sample cell window:
1
Drain liquid from module.
CAUTION
Wear appropriate safety equipment for the suspension fluids being handled.
2
Open the Optical Bench door and remove the sample cell.
3
Drain remaining liquid in the cell by disconnecting the two hoses from the line 1 and line 2 ports
on the sample cell front.
4
Loosen the two nylon screws to remove each window (
).
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......