Alignment
Straightness of the Sample and Reagent Probe (from front)
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-53
10
Straightness of the Sample and Reagent Probe (from front)
Follow the steps below to verify the "straightness" (trueness) of the Sample Probe or
Reagent Probe looking at the front of the probe. The instrument status must be in
Standby
.
Step
Action
1
Verify that the Sample or the Reagent Crane Mixer/Paddle is straight.
(Refer to Callable Diagnostics, "Sample Stir Motor and Reagent Stir
Motor" in this chapter.)
2
Rotate the Crane Assembly so that the crane assembly is facing towards
the front of the instrument.
3
Raise the probe guard (1) to expose the probes and verify that the probe is
aligned with the Crane Mixer/Paddle (2). (Refer to Figure 10.23.)
If the probe is…
then…
aligned with the Crane
Mixer/Paddle,
proceed to "Straightness of Sample and/
or Reagent Probe (from the side)."
not
aligned with the Crane
Mixer/Paddle,
adjust the probe by gently bending it to
align with the Crane Mixer/Paddle.
Figure 10.23
A014072P.EPS
2
1