Displaying/Deleting Reagent Parameters
Introduction
IMMAGE 800 Operations Manual A11403
Reagents/Calibration
September 2016
Page 6-7
6
Displaying/Deleting Reagent Parameters
Introduction
Reagent parameters may be displayed and/or deleted from the Reagent Summary.
Displaying Reagent Parameters
The instrument status must be in
Standby
in order to proceed with the steps below to
display reagent parameters.
Figure 6.3
Reagent Summary Dialog Box
Step
Action
1
Select
Rgts/Cal
from the menu bar.
2
Select
Reagent Summary [F2]
3
Select a button beside the Chem lot to be displayed.
THEN
Select the
<Display>
button to view the reagent lot parameters.
OR
Select
<Go To>
and type the Chem and Lot number.
4
Select
<OK>
from the Reagent Lot Parameters dialog box when finished
viewing the reagent lot parameters.
E010239S.EPS