Loading/Clearing UDR Buffer and Diluent
Introduction
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-90
September 2016
Loading/Clearing UDR Buffer and Diluent
Introduction
Beckman Coulter IMMAGE 800 buffers and diluents, or user-prepared solutions, may
be used as UDR buffer and diluent. However, the UDR buffer and diluent are given a
specific name on the sample or reagent carousel. The positions and lot numbers of
UDR buffer and diluent must be entered into the computer.
• Up to 4 bottles of UDR buffers may be placed on the inner section of the reagent
carousel.
• Up to 4 bottles of UDR sample diluents may be placed on the inner section of the
sample carousel.
Checking UDR Buffer/Diluent Status
The instrument status must be in
Standby
in order to proceed with the steps below to
check the buffer and diluent status before a run.
Figure 5.38
Buffer/Diluent Status Dialog Box
Step
Action
1
Select
Reagents/Cal
from the menu bar.
2
Select
Buffer/Diluent [F3]
3
Check the
% Remaining
for a sufficient amount to complete a run.
• The designation for a UDR buffer is BUF 1-4 and BUF 10-15. BUF 1-4
is designated for system buffer. BUF 10-15 is designated for user-
defined buffer.
• The designation for a UDR diluent is DIL 1-4 and DIL 10-15. DIL 1-4 is
designated for system diluent. DIL 10-15 is designated for user-defined
diluent.
E014048S.EPS