Defining a UDR Chemistry
Defining UDR Sample/Reagent Volumes
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-80
September 2016
Refer to follow the steps to enter an AGXS flagging rate value.
Defining UDR Sample/Reagent Volumes
The minimum total cuvette volume of reagent(s) and sample is 195 µL.
The maximum total cuvette volume of reagent(s) and sample is 365 µL.
The instrument status must be in
Standby
in order to proceed with the steps below to
define UDR sample and reagent volumes. Refer to Figure 5.33
Step
Action
1
From the User-Defined Chemistries screen, select the chemistry for AGXS
flagging. Refer to Figure 5.28.
2
Select
Define Edit
.
3
A user defined warning message appears. Read the message and select
<OK>
.
4
Select the
AGXS Enable
box.
5
Enter Rate Units into AGXS Limit field.
6
Select
Save [F9]
.
7
If your password is protected, then refer to the Password Setup Procedure
in this chapter and enter your user name and password.
8
Select
<OK>
.
Step
Action
1
From Page 2 of the Define/Edit User-Defined Chemistry screen, select the
Buffer
options button beside the Buffer field.
2
Select the buffer type from the list.
3
Select the
Diluent
options
<
▼
>
button beside the Diluent field.
4
Select the diluent type from the list.
(1 of 3)