Host Communications Setup
Setting Computer Port Parameters
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-51
5
Setting Computer Port Parameters
The communications port parameters must be set to match those of the LIS. The
instrument status must be in
Standby
in order to proceed with the steps below to set the
computer port parameters for communications of the IMMAGE 800 to a host
computer.
Step
Action
1
From the Setup screen, select
<10> Host Communications
. (Refer to
2
Select the options button
<
▼
>
next to one of the following parameters
• Baud Rate
• Stop Bits
• Parity
• Data Bits
3
Select the choice that matches the parameter of the host.
OR
Select
<Cancel>
to exit the dialog box without setting a parameter. The
choices are shown in Table 5.5.
4
Repeat Steps 2-3 to set additional parameters.
Table 5.5
Computer Port Parameter Choices
Baud Rate
Stop Bits
Parity
Data Bits
<1> 2400
<1> 1
<1> None
<1> 8
<2> 4800
<2> 2
<2> Even
<2> 7
<3> 9600
<3> Odd
<4> 19200