PN 623112BA
3-5
SPECIFICATIONS
PERFORMANCE SPECIFICATIONS
3
3.3
PERFORMANCE SPECIFICATIONS
Accuracy and Precision
IMMUNOPREP Reagent System Dispensing
Pipetting Module - Reagents
Pipetting Module - Specimens***
Flow-Count Fluorospheres Dispensing
* Accuracy specifications are based on the mean weight of water based on 10 replicates
(n = 10).
** Coefficient of variation is based on the mean and standard deviation of 10 replicates
(n = 10).
*** Performance claims for specimen volumes not specified above have not been established.
Note
: The performance characteristics stated in the specifications were not established by any
in vitro diagnostic assays. It is the responsibility of each laboratory to verify that the system
works for its applications.
Reagent
Volume (µL)
Precision (CV)
Accuracy (± % of volume)
IMMUNOPREP A
600
<5%
±5%
IMMUNOPREP B
265
<5%
±5%
IMMUNOPREP C
100
<5%
±5%
Dispense Volume (µL)
*Accuracy (g of water)
**Precision (CV)
1000
0.950 - 1.050
<5%
200
0.190 - 0.210
<5%
100
0.095 - 0.105
<5%
50
0.047 - 0.053
<5%
20
0.018 - 0.022
<10%
10
0.009 - 0.012
<10%
5
0.003 - 0.008
<25%
Dispense Volume (µL)
*Accuracy (g of water)
**Precision (CV)
200
0.190 - 0.210
<5%
100
0.095 - 0.105
<5%
50
0.047 - 0.053
<5%
20
0.018 - 0.022
<10%
Dispense Volume (µL)
*Accuracy (g of water)
**Precision (CV)
100
0.095 - 0.105
<5%
Summary of Contents for FP 1000
Page 4: ...PN 623112BA iv REVISION STATUS ...
Page 12: ...PN 623112BA xii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Page 30: ...PN 623112BA 3 6 SPECIFICATIONS PERFORMANCE SPECIFICATIONS ...
Page 32: ...PN 623112BA A 2 LOG SHEETS ...
Page 34: ...PN 623112BA A 4 LOG SHEETS ...
Page 36: ...PN 623112BA A 6 LOG SHEETS ...
Page 38: ...PN 623112BA A 8 LOG SHEETS ...
Page 48: ...PN 623112BA INDEX 6 INDEX ...
Page 50: ......